HU Zheng-guo, GUO Zhong-yan, ZHAN Wen-long, XIAO Guo-qing, XU Hu-shan, WANG Meng, MAO Rui-shi, Li Jia-xing, SUN Zhi-yu, CHEN Zhi-qiang, CHEN Li-xin, BAI Jie, LI Chen, ZHANG Jin-xia, LI Cun-fan. A Large Area PPAC for Using in Tumor Therapy with Heavy Ions[J]. Nuclear Physics Review, 2004, 21(1): 17-20. doi: 10.11804/NuclPhysRev.21.01.017
Citation: HU Zheng-guo, GUO Zhong-yan, ZHAN Wen-long, XIAO Guo-qing, XU Hu-shan, WANG Meng, MAO Rui-shi, Li Jia-xing, SUN Zhi-yu, CHEN Zhi-qiang, CHEN Li-xin, BAI Jie, LI Chen, ZHANG Jin-xia, LI Cun-fan. A Large Area PPAC for Using in Tumor Therapy with Heavy Ions[J]. Nuclear Physics Review, 2004, 21(1): 17-20. doi: 10.11804/NuclPhysRev.21.01.017

A Large Area PPAC for Using in Tumor Therapy with Heavy Ions

doi: 10.11804/NuclPhysRev.21.01.017
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2004-03-20
  • he principle of the heavyion therapy is introduced in this article. In order to precisely determine the irradiation profiles and positions of scanned heavyion beams, we developed a large area parallel plate avalanche counter(PPAC). The detector consists of one anode, two cathodes and two windows with 250 mm×250 mm active area . It was measured at an anode voltage of 700 V and with flowing C3F8 gas at the pressure of 700 Pa. A position resolution of 0.54 mm by correction was obtained with an αsource. As compared with the similar devices, we consider the position resolution of the PPAC can satisfy the requirement of the facility of heavyion therapy.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(1938) PDF downloads(703) Cited by()

Proportional views

A Large Area PPAC for Using in Tumor Therapy with Heavy Ions

doi: 10.11804/NuclPhysRev.21.01.017

Abstract: he principle of the heavyion therapy is introduced in this article. In order to precisely determine the irradiation profiles and positions of scanned heavyion beams, we developed a large area parallel plate avalanche counter(PPAC). The detector consists of one anode, two cathodes and two windows with 250 mm×250 mm active area . It was measured at an anode voltage of 700 V and with flowing C3F8 gas at the pressure of 700 Pa. A position resolution of 0.54 mm by correction was obtained with an αsource. As compared with the similar devices, we consider the position resolution of the PPAC can satisfy the requirement of the facility of heavyion therapy.

HU Zheng-guo, GUO Zhong-yan, ZHAN Wen-long, XIAO Guo-qing, XU Hu-shan, WANG Meng, MAO Rui-shi, Li Jia-xing, SUN Zhi-yu, CHEN Zhi-qiang, CHEN Li-xin, BAI Jie, LI Chen, ZHANG Jin-xia, LI Cun-fan. A Large Area PPAC for Using in Tumor Therapy with Heavy Ions[J]. Nuclear Physics Review, 2004, 21(1): 17-20. doi: 10.11804/NuclPhysRev.21.01.017
Citation: HU Zheng-guo, GUO Zhong-yan, ZHAN Wen-long, XIAO Guo-qing, XU Hu-shan, WANG Meng, MAO Rui-shi, Li Jia-xing, SUN Zhi-yu, CHEN Zhi-qiang, CHEN Li-xin, BAI Jie, LI Chen, ZHANG Jin-xia, LI Cun-fan. A Large Area PPAC for Using in Tumor Therapy with Heavy Ions[J]. Nuclear Physics Review, 2004, 21(1): 17-20. doi: 10.11804/NuclPhysRev.21.01.017

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return