Advanced Search

LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. doi: 10.11804/NuclPhysRev.22.01.127
Citation: LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. doi: 10.11804/NuclPhysRev.22.01.127

Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container

doi: 10.11804/NuclPhysRev.22.01.127
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2005-03-20
  • Based on the concept of relative detection efficiency, some experiments have been conducted to measure the isotopic abundance and the age of the plutonium material in a sealed container by a high-resolution γ-ray spectrometer. The PC/FRAM software analyzes the spectra. The study shows that the analysis precision of the isotopic abundance and the age will rise as the spectral measurement precision increases, and that If the spectral measurement precision is not enough, the analyzed results by the PC/FRAM software might be lower. Hence, the study suggests that a spectral peak is monitored to show lights on the spectral measurement precision during the process of spectrum acquirement.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(1748) PDF downloads(617) Cited by()

Proportional views

Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container

doi: 10.11804/NuclPhysRev.22.01.127

Abstract: Based on the concept of relative detection efficiency, some experiments have been conducted to measure the isotopic abundance and the age of the plutonium material in a sealed container by a high-resolution γ-ray spectrometer. The PC/FRAM software analyzes the spectra. The study shows that the analysis precision of the isotopic abundance and the age will rise as the spectral measurement precision increases, and that If the spectral measurement precision is not enough, the analyzed results by the PC/FRAM software might be lower. Hence, the study suggests that a spectral peak is monitored to show lights on the spectral measurement precision during the process of spectrum acquirement.

LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. doi: 10.11804/NuclPhysRev.22.01.127
Citation: LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. doi: 10.11804/NuclPhysRev.22.01.127

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return