Research of γray Measurement with Liquid Scintillation Detector
doi: 10.11804/NuclPhysRev.26.02.136
- Received Date: 1900-01-01
- Rev Recd Date: 1900-01-01
- Publish Date: 2009-06-20
-
Key words:
- liquid scintillation detector /
- light output function /
- γray energy spectrum /
- SANDII iterative method
Abstract: The light output of the BC501A liquid scintillation detector for γrays was calibrated by serial monoenergetic γray sources. The Monte Carlo simulation method to confirm Compton edge was described briefly. The result was compared with that obtained by the semiheight method. The γray esponse matrix of BC501A liquid scintillation detector was calculated by the Monte Carlo code PHRESP from PTB. The method of using response function matrix and recoil electron spectrum to unfold γray energy spectrum was introduced and the error of unfolding spectrum was also briefly discussed.