DONG Li-yuan, QI Hui-rong, LI Yu-hong, LÜ Xin-yu, OU Yanqun, CHEN Yuan-bo. Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector[J]. Nuclear Physics Review, 2012, 29(4): 379-383. doi: 10.11804/NuclPhysRev.29.04.379
Citation: DONG Li-yuan, QI Hui-rong, LI Yu-hong, LÜ Xin-yu, OU Yanqun, CHEN Yuan-bo. Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector[J]. Nuclear Physics Review, 2012, 29(4): 379-383. doi: 10.11804/NuclPhysRev.29.04.379

Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector

doi: 10.11804/NuclPhysRev.29.04.379
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2012-12-20
  • With the application of the twodimensional GEM gaseous detector in X ray imaging, the correction method of gain uniformity caused by tripleGEM avalanche structures and electric field uniformity should be studied. The paper reported the study of the tripleGEM detector with effective area 100 mm×100 mm used the Pad’s size of 9.5 mm×9.5 mm. In the test, 100 readout channels were designed. Results showed that gain remained stable over time; as air flow increases, gain from increases obviously to changes very little. Particularly, triple GEM’s gain uniformity was very good (more than 80%) and the range of energy resolution was from 0.18 to 0.22. To improve gain consistency of results, the difference value revised was obtained to be about 0.1 by the least square method. It provided a better method to improve gain uniformity of GEM detector.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(2083) PDF downloads(1123) Cited by()

Proportional views

Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector

doi: 10.11804/NuclPhysRev.29.04.379

Abstract: With the application of the twodimensional GEM gaseous detector in X ray imaging, the correction method of gain uniformity caused by tripleGEM avalanche structures and electric field uniformity should be studied. The paper reported the study of the tripleGEM detector with effective area 100 mm×100 mm used the Pad’s size of 9.5 mm×9.5 mm. In the test, 100 readout channels were designed. Results showed that gain remained stable over time; as air flow increases, gain from increases obviously to changes very little. Particularly, triple GEM’s gain uniformity was very good (more than 80%) and the range of energy resolution was from 0.18 to 0.22. To improve gain consistency of results, the difference value revised was obtained to be about 0.1 by the least square method. It provided a better method to improve gain uniformity of GEM detector.

DONG Li-yuan, QI Hui-rong, LI Yu-hong, LÜ Xin-yu, OU Yanqun, CHEN Yuan-bo. Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector[J]. Nuclear Physics Review, 2012, 29(4): 379-383. doi: 10.11804/NuclPhysRev.29.04.379
Citation: DONG Li-yuan, QI Hui-rong, LI Yu-hong, LÜ Xin-yu, OU Yanqun, CHEN Yuan-bo. Gain Uniformity Experimental Study Performed on Triple GEM Gas Detector[J]. Nuclear Physics Review, 2012, 29(4): 379-383. doi: 10.11804/NuclPhysRev.29.04.379