Efficiency Calibration Method of NaI Detector in 23.8 MeV for γ-Ray
doi: 10.11804/NuclPhysRev.32.04.440
- Received Date: 2015-01-28
- Accepted Date: 2015-12-20
- Rev Recd Date: 2015-04-14
- Publish Date: 2015-12-20
-
Key words:
- NaI(Tl) Detctor /
- energy /
- Gamma-ray /
- calibration of efficiency /
- MCNP-4C /
Abstract: In the study of D(d,γ)4He radiative capture reactions research, which is very important for fusion and astrophysics, the efficiency calibration of the detector is necessary. Due to the restriction on energy of the standard gamma source, the efficiency calibration of high-energy gamma-rays can not be calibrated by a standard source. In this paper, the method combining the experimental measurements and calculations for the efficiency calibrations of high energy gamma rays is given in the experiment of 6.13MeV gamma rays efficiency calibration using NaI detector, the efficiency curve of which is calculated by MCNP-4C and corrected experimentally. In this case, the detection efficiency of 23.8MeV gamma ray is known. The method provides a reference to high energy gamma ray efficiency calibration.
Citation: | SU Xiaobin, LIU Yang, HOU Long, WANG Zhaohui, WANG Qi. Efficiency Calibration Method of NaI Detector in 23.8 MeV for γ-Ray[J]. Nuclear Physics Review, 2015, 32(4): 440-445. doi: 10.11804/NuclPhysRev.32.04.440 |