LIU Chuanfeng, ZHOU Xiaojuan, ZHOU Jianrong, LU Liming, ZHU Lin, HE Cong, TAN Yingying, XU Hong, XIE Yuguang, WANG Xiaodong, WU Jinjie, CHEN Yuanbo, SUN Zhijia. Research on the X-ray Spot Tracing and the Position Resolution of THGEM Detector[J]. Nuclear Physics Review, 2018, 35(1): 61-65. doi: 10.11804/NuclPhysRev.35.01.061
Citation:
|
LIU Chuanfeng, ZHOU Xiaojuan, ZHOU Jianrong, LU Liming, ZHU Lin, HE Cong, TAN Yingying, XU Hong, XIE Yuguang, WANG Xiaodong, WU Jinjie, CHEN Yuanbo, SUN Zhijia. Research on the X-ray Spot Tracing and the Position Resolution of THGEM Detector[J]. Nuclear Physics Review, 2018, 35(1): 61-65. doi: 10.11804/NuclPhysRev.35.01.061
|
Research on the X-ray Spot Tracing and the Position Resolution of THGEM Detector
-
LIU Chuanfeng1,2,3
,
-
ZHOU Xiaojuan1,3
,
-
ZHOU Jianrong1,3
,
-
LU Liming1,3,5
,
-
ZHU Lin1,3,5
,
-
HE Cong1,3,5
,
-
TAN Yingying1,3,5
,
-
XU Hong1,3
,
-
XIE Yuguang1,4
,
-
WANG Xiaodong5
,
-
WU Jinjie6
,
-
CHEN Yuanbo1,3,4
,
-
SUN Zhijia1,3,4
,
- 1.
Institute of High Energy Physics Chinese Academy of Sciences, Beijing 10049, China;
- 2.
Centre of Excellence for Advanced Materials, Dongguan 523808, Guangdong, China;
- 3.
Dongguan Neutron Science Center, Dongguan 523803, Guangdong, China;
- 4.
State Key Laboratory of Particle Detection and Electronics, Beijing 100049, China;
- 5.
School of Nuclear Science and Technology, University of South China, Hengyang 421000, Hunan, China;
- 6.
National Institute of Metrology, Beijing 100029, China
Funds:
National Key R&D Program of China (2017YFA0403702); National Natural Science Foundation of China (11635012, 11775243); Guangdong Innovative and Entrepreneurial Research Team Program (2016ZT06G025); Scientific Instrument Developing Project of Chinese Academy of ScienceS(YZ201512)
- Received Date: 2017-11-27
- Rev Recd Date:
2018-01-19
- Publish Date:
2018-03-20
-
Abstract
This paper introduces the detector of X-ray spot tracing based on the THGEM. It is used for X ray single photon measurement technology research and measurement with an effective area of 200 mm×200 mm. The circuit board has 128 array of high speed readout electronics based on ASIC and FPGA. And it is divided into two parts:the middle area of high sensitive and the peripheral area of low sensitive. The two parts share a set of readout electronics. The experiment tests the position and the fine structure of the beam line produced by X ray machine in National Institute of Metrology. The results of the experimental show that the detector realizes both the spot tracing and the fine structure of the spot. The position resolution is 0.63 mm in x direction and 0.62 mm in y direction in high sensitive of the detector, which achieve the desired goal of preliminary design.
-
References
[1]
|
SAULI F. Nucl Instr Meth A, 1997, 386:531. |
[2]
|
BACHMANN S, BRESSAN A, KAPPLER S, et al. Nucl Instr & Meth A, 2001, 471:971. |
[3]
|
ALEXEEV M, BIRSA R, BRADAMANTE F, et al. Nucl Instr & Meth A, 2009, 612(1):174. |
[4]
|
BADERTSCHER A, KNECHT L, LAFFRANCHI M, et al. Nucl Instr & Meth A, 2010, 617:188. |
[5]
|
WANG Yanfeng, SUN Zhijia, ZHOU Jianrong, et al. SCIENCE CHINA Physics, Mechanics & Astronomy, 2013, 56(10):1897. |
[6]
|
WANG Tuo, ZHOU Jianrong, SUN Zhijia, et al. Nuclear Physics Review, 2014, 31(1):69. (in Chinese) (王拓, 周建荣, 孙志嘉, 等. 原子核物理评论, 2014, 31(1):69.) |
[7]
|
CHECHIK R, BRESKIN A, SHALEM C, et al. Nucl Instr & Meth A, 2004, 535:303. |
[8]
|
ZHENG Xiaocui, LI Yulan, LAI Yongfang, et al. Nuclear Electronics & Detection Technoogy, 2007, 27(1):89. (in Chinese) (邓晓翠, 李玉兰, 来永芳, 等. 核电子学与探测技术, 2007, 27(1):89.) |
[9]
|
LI Ke. Experiment Study of Thermal Neutron Detector Based on nTHGEM[D]. Lanzhou:Lanzhou University, 2016:20. (in Chinese) (李科. 基于nTHGEM热中子探测器的实验研究[D]. 兰州:兰州大学, 2016:20.) |
[10]
|
The ROOT Team. The ROOT users guide[M/OL].[2017-12-01]. http://root.cern.ch/Download/doc/User/doc/User_Guide_526.pdf. |
-
-