Advanced Search

LIU Ling, QI Huirong, DAI Hongliang, WEN Zhiwen, YUAN Zhiyang, WANG Haiyun, ZHANG Jian, OUYANG Qun, SHAO Jianxiong. Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source[J]. Nuclear Physics Review, 2019, 36(3): 351-356. doi: 10.11804/NuclPhysRev.36.03.351
Citation: LIU Ling, QI Huirong, DAI Hongliang, WEN Zhiwen, YUAN Zhiyang, WANG Haiyun, ZHANG Jian, OUYANG Qun, SHAO Jianxiong. Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source[J]. Nuclear Physics Review, 2019, 36(3): 351-356. doi: 10.11804/NuclPhysRev.36.03.351

Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source

doi: 10.11804/NuclPhysRev.36.03.351
Funds:  National Natural Science Fundation of China(11675197, 11775242)
  • Received Date: 2019-02-21
  • Rev Recd Date: 2019-03-27
  • Publish Date: 2019-09-20
  • Aimed to the Multipurpose Reflectance spectrometer device(MR) physics requirements, which is that the active area with 200 mm×200 mm, the detection efficiency with more than 50% and the position resolution with better than 2 mm, for China Spallation Neutron Source (CSNS) in Dongguang, China. A 3He two-dimensional multi-wire proportional chamber position sensitive neutron detector has been developed by the Institute of High Energy Physics of the Chinese Academy of Sciences (IHEP, CAS). The high-pressure two-dimensional multiwire proportional chamber neutron detector filled with 3He/C3H8(6 atm/2.5 atm) mixture gases and the multiwire technology. Firstly, the detection uniformity was tested using the Am/Be neutron source in IHEP and the uniformity could reach to 95.1%. After assembled at #20 beam station in CSNS, the performance of the detector were studied under the neutron beam. Under the selection neutron beam with 2.8Å by the time of flight (TOF), the two-dimensional position resolution and two-dimensional imaging of the detector were measured. The position resolution is 0.99 mm in the X direction (perpendicular to the direction of the anode wire), 1.36 mm in the Y direction (parallel to the direction of the anode wire) and the two-dimensional imaging is clear to identity. The results show that the detector has a good resolution and two-dimensional imaging capability, and it could meet the physics requirements of Multipurpose Reflectance spectrometer device in CSNS.
  • [1] WANG Fangwei, LIANG Tianjiao, YIN Wen, et al. SCIENCE CHINA Physics, Mechanics Astronomy, 2013, 56(12):2410.
    [2] ZUO Taisen, CHENG Hesheng, CHEN Yuanbo, et al. Chinese Physics C, 2016, 40(7):076204.
    [3] TAN Zhixin, TANG Jingyu, JING Hantao, et al. Nucl Instr and Meth A, 2018, 889:122.
    [4] YU Quanzhi, YIN Wen, LIANG Tianjiao. Acta Phys Sin, 2010, 60(5):052501. (in Chinese) (于全芝, 殷文, 梁天骄. 物理学报, 2010, 60(5):052501.)
    [5] WEI Jie, CHEN Hesheng, CHEN Yuanbo, et al. Nucl Instr and Meth A, 2009, 600(1):10.
    [6] ZHAN Xiaozhi, XIAO Songwen, WU Yanyan, et al. Mod Phys, 2016, 58(S1):023. (in Chinese) (詹晓芝, 肖松文, 吴岩延. 现代物理知识, 2016, 58(增刊1):023.)
    [7] CHEN Hesheng. Mod Phys, 2016, 2016(01):03. (in Chinese) (陈和生. 现代物理知识, 2016, 2016(01):03.)
    [8] SHEN Fei, LIANG Tai Ran, YIN Wen, et al. Acta Phys Sin, 2014, 58(15):152801. (in Chinese) (沈飞, 梁泰然, 殷雯, 等. 物理学报, 2014, 58(15):152801.)
    [9] Zhang Liying, JING Hantao, TANG Jingyu, et al. Appl Radiat Isot 2018, 132:212.
    [10] CHEN Hesheng, WANG Xunli. Nature Materials, 2016, 15:689.
    [11] NISHIYAMA Y, LANGAN P, CHANZY H. J Am Chem Soc, 2002, 124:9074.
    [12] LI Tianfu, CHEN Dongfeng, WANG Hongli, et al. Acta Phys Sin, 2009, 58(11):7993. (in Chinese) (李天富, 陈东风, 王洪立, 等. 物理学报, 2009, 58(11):7993)
    [13] Huang Chaoqiang, CHEN Bo, LI Xinxi, et al. Acta Phys Sin, 2008, 57(1), 364. (in Chinese) (黄朝强, 陈波, 李新喜, 等. 物理学报, 2008, 57(1):364.)
    [14] WEN Zhiwen, QI Huirong, ZHANG Yulian, et al. Acta Phys sin, 2018, 67:072901. (in Chinese) (温志文, 祁辉荣, 张余炼, 等. 物理学报, 2018, 67(7):072901.)
    [15] WENK H R, LUTTERTTI, L, VOGEL, S. Nucl Instr and Meth A, 2003, 515(3):575.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(1500) PDF downloads(59) Cited by()

Proportional views

Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source

doi: 10.11804/NuclPhysRev.36.03.351
Funds:  National Natural Science Fundation of China(11675197, 11775242)

Abstract: Aimed to the Multipurpose Reflectance spectrometer device(MR) physics requirements, which is that the active area with 200 mm×200 mm, the detection efficiency with more than 50% and the position resolution with better than 2 mm, for China Spallation Neutron Source (CSNS) in Dongguang, China. A 3He two-dimensional multi-wire proportional chamber position sensitive neutron detector has been developed by the Institute of High Energy Physics of the Chinese Academy of Sciences (IHEP, CAS). The high-pressure two-dimensional multiwire proportional chamber neutron detector filled with 3He/C3H8(6 atm/2.5 atm) mixture gases and the multiwire technology. Firstly, the detection uniformity was tested using the Am/Be neutron source in IHEP and the uniformity could reach to 95.1%. After assembled at #20 beam station in CSNS, the performance of the detector were studied under the neutron beam. Under the selection neutron beam with 2.8Å by the time of flight (TOF), the two-dimensional position resolution and two-dimensional imaging of the detector were measured. The position resolution is 0.99 mm in the X direction (perpendicular to the direction of the anode wire), 1.36 mm in the Y direction (parallel to the direction of the anode wire) and the two-dimensional imaging is clear to identity. The results show that the detector has a good resolution and two-dimensional imaging capability, and it could meet the physics requirements of Multipurpose Reflectance spectrometer device in CSNS.

LIU Ling, QI Huirong, DAI Hongliang, WEN Zhiwen, YUAN Zhiyang, WANG Haiyun, ZHANG Jian, OUYANG Qun, SHAO Jianxiong. Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source[J]. Nuclear Physics Review, 2019, 36(3): 351-356. doi: 10.11804/NuclPhysRev.36.03.351
Citation: LIU Ling, QI Huirong, DAI Hongliang, WEN Zhiwen, YUAN Zhiyang, WANG Haiyun, ZHANG Jian, OUYANG Qun, SHAO Jianxiong. Experimental Study on Two-dimensional Position Sensitive Detector Filled with Helium-3 for the China Spallation Neutron Source[J]. Nuclear Physics Review, 2019, 36(3): 351-356. doi: 10.11804/NuclPhysRev.36.03.351
Reference (15)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return