Design and Implementation of Simplified Readout Circuits for Multi anode Position sensitive Photomultiplier Tubes

doi: 10.11804/NuclPhysRev.27.01.063
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2010-03-20
  • The multianode position sensitive photomultiplier tube(PSPMT) has been used for the development of highresolution γ camera module. This study is aimed to investigate the simplified readout design for the multianode flat panel PSPMT. Based on the Hamamatsu H8500 PSPMT with 64 multianodes, we designed and developed two simplified readout circuits, discretized positioning circuit(DPC) and symmetric charge division circuit(SCDC). The truncated centerofgravity(TCOG) positioning method was used to optimize the readout signal processing and then further reduce the readout channels by a resistor chain. The preliminary results show that the SCDC readout with the TCOG could significantly reduce the nonlinearity of positioning and improve the effective fieldofview of the detector by about 20% as compared to the DPC readout.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(3388) PDF downloads(917) Cited by()

Proportional views

Design and Implementation of Simplified Readout Circuits for Multi anode Position sensitive Photomultiplier Tubes

doi: 10.11804/NuclPhysRev.27.01.063

Abstract: The multianode position sensitive photomultiplier tube(PSPMT) has been used for the development of highresolution γ camera module. This study is aimed to investigate the simplified readout design for the multianode flat panel PSPMT. Based on the Hamamatsu H8500 PSPMT with 64 multianodes, we designed and developed two simplified readout circuits, discretized positioning circuit(DPC) and symmetric charge division circuit(SCDC). The truncated centerofgravity(TCOG) positioning method was used to optimize the readout signal processing and then further reduce the readout channels by a resistor chain. The preliminary results show that the SCDC readout with the TCOG could significantly reduce the nonlinearity of positioning and improve the effective fieldofview of the detector by about 20% as compared to the DPC readout.

ZHAO Cui-lan, QI Yu-jin, #, SHI Liang, JIN Qing, DAI Qiu-sheng. Design and Implementation of Simplified Readout Circuits for Multi anode Position sensitive Photomultiplier Tubes[J]. Nuclear Physics Review, 2010, 27(1): 63-67. doi: 10.11804/NuclPhysRev.27.01.063
Citation: ZHAO Cui-lan, QI Yu-jin, #, SHI Liang, JIN Qing, DAI Qiu-sheng. Design and Implementation of Simplified Readout Circuits for Multi anode Position sensitive Photomultiplier Tubes[J]. Nuclear Physics Review, 2010, 27(1): 63-67. doi: 10.11804/NuclPhysRev.27.01.063

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return